Figure 1 | Journal of Hematology & Oncology

Figure 1

From: Prozone effect of serum IgE levels in a case of plasma cell leukemia

Figure 1

Immulite® readings of multiple serial dilutions of the same sample showing prozone effect. The calculated concentration of IgE based on the final measured reading times the dilution factor is plotted along the X axis against the actual instrument reading (counts per second) along the Y axis. Both axes are logarithmic and the dilutions that were used were 1:100,000, 1:10,000, 1:1000, 1:100, and undiluted. In parenthesis at each dilution point is the reading reported by the instrument. The dotted line represents the highest reported value of 2000 IU/mL. The lines connecting each point are for illustration and do not represent the actual values at intermediate dilutions.